Nitric Oxide (NO), produced endogenously from L-Arginine by nitric oxide synthatases, plays an important role in many physiological processes including vascular regulation, immune responses, and neural communication. NO is extremely unstable and undergoes repaid oxidative degradation to nitrite (NO2-) and nitrate (NO3-), which can be spectrophotometrically determined. ScienCell's Colorimetric Nitric Oxide Assay kit provides an accurate measurement of NO level in a simple two-step process: the reduction of nitrate to nitrite by vanadium (III) chloride, followed by quantification of nitrite by Griess reaction.
Kit Components
Cat. No
# of Vials
Reagents
Quantity
Storage
8098a
1
Nitrate Standard, 200 µM
2.5 ml
40C, dark
8098b
1
Vanadium Chloride
25 ml
40C, dark
8098c
1
Griess Reagent I
12.5 ml
40C, dark
8098d
1
Griess Reagent II
12.5
40C, dark
8098e
1
20× ZnSO4
12.5
40C
Quality Control
The ScienCellT Colorimetric Nitric Oxide Assay is applied to nitrate standards serially diluted from 200 to 3.13 µM. Standard curves obtained with different incubation time/temperature are shown in Figure 1.
Procedures
Deproteination of samples
Mix 285 µl of each sample (e.g. cell culture supernatant) with 15 µl of 20× ZnSO4 in a 1.5 ml micro tube, vortex for 1 minute, centrifuge at 10,000 RCF for 10 min at 4ºC, and transfer 100 µl/well of supernatant into each wells of the 96-well plate. We recommend that you prepare three replicates for each sample.
Preparation of nitrate standards
Obtain 8 test tubes and label them A through H. Add 300 µl of DI H2O into tubes B through H.
Add 300 µl of the 200 µM Nitrate Standard solution into tube A.
Add 300 µl of the 200 µM Nitrate Standard solution into tube B and mix well to get the 100 µM nitrate standard.
Transfer 300 µl of the 100 µM Nitrate standard from tube B to tube C and mix well to get the 50 µM nitrate standard.
Repeat step 3 for tubes D-G to serially dilute the nitrate standards. Do not add any nitrate solution to tube H, which serves as the blank.
Obtain a 96-well test plate; prepare 3 replicates of each nitrate standard by aliquoting 100 µl/well of each nitrate standard into triplicate wells of the 96-well plate, according to the plate format shown in Table 1.
Measurement of NO3-/NO2-
Make fresh reaction "cocktail" by mixing 100 µl of vanadium chloride with 50 µl of Griess reagent I and 50 µl of Griess reagent II for each well of 96-well plate. Prepare adequate reaction "cocktail" based on the number of samples/standards to be assessed.
Add 200 µl of reaction "cocktail" to each well containing 100 µl of sample or nitrate standard and incubate for 30-120 min at room temperature, protected from light*. Solutions should turn a pale pink color.
Measure the absorbance on an ELISA plate reader with a test wavelength at 540 nm and a reference wavelength at 630 nm, and subtract the 630 nm background absorbance from the 540 nm measurement.
Calculation
Average the calibrated absorbance values (OD540nm) of each sample, nitrate standard and blank wells.
Subtract the average OD540nm of blank from the average OD540nm of each sample and nitrate standard.
Generate the standard curve by plotting the calibrated OD540nm of the nitrate standards against the nitrate concentrations, as shown in Figure 1.
Determine the total concentration of nitrate and nitrite of each sample based on the standard curve.
*The time of incubation depends on the NO3-/NO2- concentration of the samples. Dilute the samples if they are too concentrated. The color could get lost if the reaction goes too far due to too long incubation time or too concentrated samples. The time of incubation can be shortened by incubating at 37ºC instead of room temperature.
Click the picture for large image
Fig A
Fig B
The ScienCellT Colorimetric Nitric Oxide assay is applied to nitrate standards serially diluted from 200 to 3.13 µM. Standard curves obtained with incubation time of 30, 60 and 120 min at room temperature (A) and 37ºC (B) are compared.